6.7 Subwindow Sensors
■■■■■■■■■■■■■■■■■■■■■■
1302
■■■■■■■■
tiamo 2.2
for ISE (+3)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
23.0 mV
for ISE (-3)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-16.0 mV
for ISE (+4)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
17.0 mV
for ISE (-4)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-12.0 mV
pH(0) monitoring (for pH electrodes only)
on | off (Default value: off)
If this option is enabled then a check will be made to see whether the elec-
trode zero point pH(0) determined during the calibration of a pH electrode
is within the defined limits. If it is outside them, a corresponding message
appears and you can select whether you will nevertheless accept the value
or whether the previous value is to be retained.
Lower limit
Lower limit for pH(0).
Range
-2.147E+9 ... 2.147E+9
Default value
6,750
Upper limit
Upper limit for pH(0).
Range
-2.147E+9 ... 2.147E+9
Default value
7,250
E(0) monitoring (for ISE electrodes only)
on | off (Default value: off)
If this option is enabled then a check will be made to see whether the elec-
trode zero point E(0) determined during the calibration of an ISE electrode
is within the defined limits. If it is outside them, a corresponding message
appears and you can select whether you will nevertheless accept the value
or whether the previous value is to be retained.