DET U - Measured value window

Dialog window: Method

DET U Properties... Potentiometric evalua-

tion

Evaluation with measured value window (U) [New]/[Proper-

ties]

Measured value window#

Measured value windows are regions on the measured value axis for which
different parameters for potentiometric evaluation can be defined. Only
those endpoints which lie within these windows and additionally meet the
parameters defined for each window will be recognized. Only one endpoint
will be recognized per window.

Lower limit

Lower limit for the measured value window.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

-2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

-2000 mV

Upper limit

Upper limit for the measured value window.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

2000 mV

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.

Titrando, 855
Range

0 ... 200

Default value

5

EP recognition

Filter for the recognition of equivalence points: