6 Configuration
tiamo 2.2
■■■■■■■■
1219
Comment on modification of sample data (live)
If this check box is activated then each time that a determination is modified
a modification reason and a modification comment must be entered; these
are saved in the determination and shown in the database in the subwindow
Information on the tab Sample. The reason and comments are also recor-
ded in the Audit Trail.
Selection
on | off
Default value
off
FDA default: on
6.2.2.5
Signatures
Tab: Configuration
▶ Tools ▶ Security settings ▶ Signatures
On the tab Signatures the parameters for the electronic signatures can be
configured.
Settings according to 21 CFR 11
[Set]
Sets the parameters to default settings for FDA-compliant work.
Inactivity delay
If this check box is activated then the dialog window for signing will be
automatically closed when the entered time limit has expired.
Selection
on | off
Default value
off
FDA default: on
Range
1 ... 60 min
Default value
10 min
FDA default: 10 min
Remove password after signature
If this check box is activated then the password must be entered again after
each signature.
Selection
on | off
Default value
off
FDA default: on
Sign each determination separately
If this check box is activated then each determination contained in the
determination overview must be signed individually.
Selection
on | off
Default value
off
FDA default: on