7 properties - glp – Metrohm tiamo 2.2 Manual User Manual
Page 1447
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
tiamo 2.2
■■■■■■■■
1433
With the 774 Oven Sample Processor display of the initial temperature set
on the device only.
Range
50 ... 250 °C
Default value
off °C
Max. temperature
With the 774 Oven Sample Processor display of the maximum temperature
set on the device only.
With the 774 Oven Sample Processor display only.
Range
50 ... 250 °C
Default value
250 °C
Temperature correction
Enter the temperature correction. This parameter directly affects the tem-
perature control and allows it to be adjusted. In this way any temperature
difference that may occur between the oven and sample temperature can
be compensated.
With the 774 Oven Sample Processor display of the temperature correction
set on the device only.
Range
-10 ... +10 °C
Default value
0 °C
7.14.7
Properties - GLP
Tab: Configuration
▶ Device ▶ [Edit] ▶ Properties... ▶ 'Properties - 'Device type' -
'Device name'
▶ GLP
GLP test date
Date of the last GLP test. This date can be selected by clicking on
in the
dialog window Select date (see Chapter 2.5.1, page 83).
Comment GLP test
Comment on GLP test.
Input
1000 characters
Monitoring of GLP validity
Monitoring of GLP validity
on | off (Default value: off)
If this option is switched on then the time interval for the GLP test will be
monitored.