6 Configuration
tiamo 2.2
■■■■■■■■
1303
Lower limit
Lower limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-2000.0 mV
Upper limit
Upper limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
2000.0 mV
Cell constant monitoring (only for conductivity sensors)
on | off (Default value: off)
If this option is enabled, then a check will be made to see whether the value
determined during the determination of the cell constant of a conductivity
sensor is within the defined limits. If it is outside them, a corresponding
message appears and you can select whether you will nevertheless accept
the value or whether the previous value is to be retained.
Lower limit
Lower limit for cell constant.
Range
0.001 ... 500 /cm
Default value
0.001 /cm
Upper limit
Upper limit for cell constant.
Range
0.001 ... 500 /cm
Default value
0.001 /cm
6.7.3.5
Properties - History
Tab: Configuration
▶ Sensors ▶ Edit ▶ Sensor ▶ History
Note
This tab only appears for pH and ISE electrodes.
History table
The table contains the last 10 calibrations for the selected sensor and cannot
be edited or sorted. The individual calibrations are arranged chronologically
according to date so that the most recent determination appears last. For
pH electrodes, the Slope and the pH(0) are displayed, for ISE electrodes the