Metrohm tiamo 2.2 Manual User Manual
Page 1022
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1008
■■■■■■■■
tiamo 2.2
Max. waiting time
If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:
Range
0 ... 999999 s
Default value
52 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Range
0.1 ... 999999.0 s
Default value
2.0 s
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous, then
the temperature will be measured continuously.
Range
-20.0 ... 150°C
Default value
25.0
Switch off stirrer during measurement
Switch off stirrer during measurement
on | off (Default value: off)
If this option is activated, then the stirrer is switched off during the meas-
urement. In this case the following two parameters can be edited:
Stir before measurement
Stirring time between end of dosing and measurement.
Range
0 ... 999999 s
Default value
0 s
Pause before measurement
Waiting time before the measurement during which no stirring takes place.
Range
0 ... 999999 s
Default value
0 s