5 Method
tiamo 2.2
■■■■■■■■
833
Titrino
Range
-170.0 ... 500.0 °C
Default value
500.0 °C
Upper hysteresis
Upper hysteresis of the temperature. When the temperature once again falls
below the upper limit by this hysteresis value, the action Temperature
upper limit ok is triggered
Titrando, 855
Range
0.0 ... 150.0 °C
Default value
0.2 °C
Action
Selection of the event to be triggered if the upper or lower limit is exceeded:
Selection
Stop determination | Stop command (only
Titrando) | Wait for [Continue] | Wait for limit
ok | none
Default value
none
Stop determination
The running STAT command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running STAT command is quit, then the next command is executed.
Wait for [Continue]
Reagent dosing in the current STAT command is interrupted and a mes-
sage appears. As soon as the monitored measured value is again within
the limits (including hysteresis) reagent dosing can be resumed by press-
ing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command is interrupted. As soon
as the monitored measured value is again within the limits (including
hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.
Track call on limit exceeding
In this table, which cannot be edited directly, a maximum of 20 entries can
be defined as to which track is to be started automatically when a particular
limit value is exceeded.
[New]
Opens the dialog window Track call # in which the parameters for the
track call can be entered.