5 Method
tiamo 2.2
■■■■■■■■
827
Stop volume
Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stops when the preset time has elapsed after the start conditions have been
fulfilled.
Range
0 ... 999999 s
Selection
off
Default value
off
since
Specifies when the stop time begins to be counted down (will not be dis-
played if the Stop time is disabled):
Selection
Start | Control point first reached | Last dosing
Default value
Start
Start
The stop time starts after the start volume has been dosed and after the
following pause time.
Control point first reached
The stop time starts as soon as the control point has been reached for
the first time.
Last dosing
The stop time starts after the last increment has been dosed, i. e. the stop
time is reset to zero again after each dosing step.
Stop rate
Stops when the current dosing rate is less than the defined stop rate. This
criterion is checked 10 s after the start at the earliest. The maximum stop
rate depends on the cylinder volume of the exchange unit or dosing unit
used.