5 Method
tiamo 2.2
■■■■■■■■
487
Window = Time
Range
0 ... 999999 s
Default value
999999 s
Gran evaluation
Note
This evaluation method is possible only with Titrando and 855.
Gran evaluation
on | off (Default value: off)
If this option is enabled, then titration curves will be evaluated in accordance
with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 878).
Initial volume
Volume which must be present in the measuring vessel before the command
is started.
Range
0.01 ... 9999.99 mL
Default value
50.00 mL
Lower limit
Value for the lower limit of the Gran evaluation range.
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Upper limit
Value for the upper limit of the Gran evaluation range.
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
5.6.3.2.3.8
DET U - Additional measured values
Tab: Method
▶ DET U ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Input
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings: