5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
936
■■■■■■■■
tiamo 2.2
Parameters
The parameters for the command MEAS T are configured on the following
4 tabs:
■
General/Hardware
Parameters for devices, sensors and stirrers.
■
Measuring parameters
Parameters for setting the measurement procedure.
■
Evaluations
Defining further methods for evaluating measurement curves.
■
Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.
Command variables
The following command variables are generated in the method run by the
command MEAS T and can be used in formulas under the designation
'Command name.Variable designation':
Identification
Description
.BP{x}.MEA
Measured value (temperature) for the break point ×
(1…9) in the unit of the measured value
.BP{x}.TEM
Temperature for the break point × (1…9) in °C
.BP{x}.TIM
Time for the break point × (1…9) in s
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.DBL
Total duration for the processing of the command in s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode zero point of the sensor used for the com-
mand (dimensionless)
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command
has never ended
.FP{x}.MEA
Measured value (temperature) for the fixed endpoint ×
(1…9) in the unit of the measured value
.FP{x}.TEM
Temperature for the fixed endpoint × (1…9) in °C