2 meas u - general/hardware, General/hardware – Metrohm tiamo 2.2 Manual User Manual
Page 913
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
899
Identification
Description
.SLO
Electrode slope of the sensor used for the command (in
mV for ISE sensors)
.STY
Type of stop with which the command was stopped:
1 = normal;0 = manual or after error
5.6.4.4.2
MEAS U - General/Hardware
Tab: Method
▶ MEAS U ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the sensor and the stirrer are defined
on this tab.
Device
Device name
Selection of a device from those available in the device table. The only devi-
ces which shall be offered are those with which the command can run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start of
the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead
If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
Device types | Titrando
Default value
Titrando
Sensor
Measuring input
Selection of the measuring input to which the sensor is connected.