3 det upol - start conditions, Start conditions – Metrohm tiamo 2.2 Manual User Manual
Page 540
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
526
■■■■■■■■
tiamo 2.2
5.6.3.2.5.3
DET Upol - Start conditions
Tab: Method
▶ DET Upol ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the titration
is started.
Initial measured value
Note
Is displayed only for Titrando and 855.
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:
Measured value drift
Drift for the measured value acceptance during the titration.
Titrando, 855
Range
0.01 ... 99.90 µA/min
Default value
20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Titrino
Range
0.05 ... 99.90 µA/min
Default value
20.00 (slow), 50.00 (optimal), 80.00 (fast) µA/
min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time