9 met ipol - measured value window – Metrohm tiamo 2.2 Manual User Manual
Page 634
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
620
■■■■■■■■
tiamo 2.2
[Properties]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880).
[Delete]
Deletes the external measured value selected in the table.
5.6.3.3.4.9
MET Ipol - Measured value window
Dialog window: Method
▶ MET Ipol ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with measured value window (Ipol) ▶ [New]/[Proper-
ties]
▶ Measured value window#
Measured value windows are regions on the measured value axis for which
different parameters for potentiometric evaluation can be defined. Only
those endpoints which lie within these windows and additionally meet the
parameters defined for each window will be recognized. Only one endpoint
will be recognized per window.
Lower limit
Lower limit for the measured value window.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
-2000 mV
Upper limit
Upper limit for the measured value window.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
2000 mV
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.
Range
1 ... 999 mV
Default value
30 mV