5 Method
tiamo 2.2
■■■■■■■■
593
Lower limit
Value for the lower limit of the Gran evaluation range.
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Upper limit
Value for the upper limit of the Gran evaluation range.
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
5.6.3.3.3.8
MET U - Additional measured values
Tab: Method
▶ MET U ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Input
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings.
Additional calculated measured values
Additional calculated measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can be
calculated from existing measured values and saved, which can also be pre-
sented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.
[New]
Opens the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be entered (see Chapter
5.6.3.11.1, page 879).
[Properties]
Opens the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be edited (see Chapter
5.6.3.11.1, page 879).
[Delete]
Deletes the calculated measured value selected in the table.