5 Method
tiamo 2.2
■■■■■■■■
829
Titrando, 855
Range
-20,000 ... 20,000
Default value
-20,000
Titrino
Range
-20.00 ... 20.00
Default value
-20.00
Lower hysteresis pH
Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value lower
limit ok is triggered.
Titrando, 855
Range
0.000 ... 20,000
Default value
0.020
Upper limit pH
Upper limit of the measured value. If the measured value exceeds this limit,
then the event Measured value upper limit exceeded is triggered.
Titrando, 855
Range
-20,000 ... 20,000
Default value
20,000
Titrino
Range
-20.00 ... 20.00
Default value
20.00
Upper hysteresis pH
The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.
Titrando, 855
Range
0.000 ... 20,000
Default value
0.020
Action
Selection of the event to be triggered if the upper or lower limit is exceeded:
Selection
Stop determination | Stop command (only
Titrando) | Wait for [Continue] | Wait for limit
ok | none
Default value
none