5 Method
tiamo 2.2
■■■■■■■■
567
Gran evaluation
Note
This evaluation method is possible only with Titrando and 855.
Gran evaluation
on | off (Default value: off)
If this option is enabled, then titration curves will be evaluated in accordance
with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 878).
Procedure
Selection of the Gran procedure.
Selection
Normalized | Standard
Default value
Normalized
Initial volume
Volume which must be present in the measuring vessel before the command
is started.
Range
0.01 ... 9999.99 mL
Default value
50.0 mL
Lower limit pH
Value for the lower limit of the Gran evaluation range.
Range
-20.000 ... 20.000
Default value
-20.000
Upper limit pH
Value for the upper limit of the Gran evaluation range.
Range
-20.000 ... 20.000
Default value
20.000
5.6.3.3.2.8
MET pH - Additional measured values
Tab: Method
▶ MET pH ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Input
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns