5 Method
tiamo 2.2
■■■■■■■■
727
Titration direction
Selection of the titration direction. If two endpoints are set, then the titration
direction is already established and the following selection of titration
direction is ignored.
Selection
+ | - | auto
Default value
auto
+
positive measured value alteration
-
negative measured value alteration
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.
Extraction time
The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). For example, it makes sense to enter an
extraction time for the titration of samples that are not easily soluble.
Titrando, 719, 720, 736, 751, 758, 785, 794, 798, 799
Range
0 ... 999999 s
Temperature
Manually entered titration temperature. If a temperature sensor is connected
and the Temperature measurement is set on the General/Hardware
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Range
0.1 ... 999999 s
Default value
2.0 s