2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

tiamo 2.2

Identifica-
tion

Description

Commands

.CAL

Status of the calibration: 1 = cali-
brated normally
, 0 = canceled,
because reference wavelength was
not found.

CAL Spec

.CBY

Command status; 1 = Condition-
ing activated
, 0 = Conditioning
deactivated

SET, KFT, KFC,
BRC

.CLC

Cell constant of the sensor used in
the command (with conductivity
sensors)

MEAS Cond, CAL
Cond

.COK

Command status; 1 = Condition
requirement fulfilled
, 0 = Con-
dition requirement not fulfilled

SET, KFT, KFC,
BRC

.CONC

Concentration of the solution used
for the command

DET, MET, SET,
KFT, STAT, ADD,
DOS, LQH,
STDADD

.CP{#}.MEA

Measured wavelength of the cali-
bration point in nm. If no measured
peak can be assigned, then not
found
will be entered.

CAL Spec

.CP{#}.REF

Reference wavelength of the cali-
bration point in nm.

CAL Spec

.CYL

Cylinder volume of the exchange or
dosing unit used for the command

DET, MET, SET,
KFT, STAT,
STDADD dos,
STDADD auto,
ADD, DOS, LQH,
PREP, EMPTY

.DBL

Total duration for the processing of
the command in s

DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS,
CAL MEAS,
STDADD, DOS

.DEF

Default value which is defined in the
configuration for an IO port

ANALOG OUT,
DIGITAL OUT

.DRI

Current and/or last drift for drift cor-
rection in µL/min

SET, KFT, KFC,
BRC