Metrohm tiamo 2.2 Manual User Manual
Page 1174
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1160
■■■■■■■■
tiamo 2.2
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead
If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
Device types | Titrando
Default value
Titrando
Remote box
Selection of a remote box attached to the device.
Titrando, 855, 856, 867, 814, 815, 864, 874
Selection
1 | 2 | 3 | 4
Default value
1
Set lines
Output signal
Input of the binary pattern for the output signal or selecting a predefined
signal pattern.
It is possible to enter the following characters:
0 = line inactive
1 = line active
* = any line status
p = set pulse (pulse length = 200 ms). If an impulse with a different length
is to be emitted, a corresponding template must be defined.
Titrando, 855, 856, 867, 751, 756, 758, 784, 785, 795, 798, 799, 831,
Sample Processor
Selection
Bit-pattern with exactly 14 characters (0, 1, *,
p) | ************** | Signal pattern
Default value
**************
702, 716, 718, 719, 720, 721, 736, 794
Selection
Bit-pattern with exactly 8 characters (0, 1, *,
p) | ******** | Signal pattern
Default value
********