5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
622
■■■■■■■■
tiamo 2.2
EP criterion mV
Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.
Range
1 ... 999
Default value
30
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.3.4.11
MET Ipol - Fix endpoint evaluation
Dialog window: Method
▶ MET Ipol ▶ Properties... ▶ Additional evaluations ▶ Fix
endpoint evaluation
▶ [New]/[Properties] ▶ Fix endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-2000.0 ... 2000.0 mV