6.6 Subwindow Titrants/Solutions
■■■■■■■■■■■■■■■■■■■■■■
1278
■■■■■■■■
tiamo 2.2
IEU
Input
8 characters
Default value
'empty'
EU
Input
10 characters
Default value
'empty'
Parameters for preparation
Configures the parameters to be used in the command PREP.
Volume
Volume to be dosed in during preparation.
Range
0.00000 ... 99999.99999 mL
Selection
Cylinder volume
Default value
Cylinder volume
Cylinder volume
The whole contents of the cylinder will be dosed in.
Cycles
Number of rinsing cycles during preparation.
Range
1 ... 9
Default value
2
Dosing rate
Speed at which the solution is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit used (see below).
When the function is carried out the speed will automatically be reduced to
the largest possible value.
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Filling rate
Speed at which the cylinder is to be filled. The maximum filling rate depends
on the cylinder volume of the exchange unit used (see below). When the
function is carried out the speed will automatically be reduced to the largest
possible value.
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum