10 stat u - fix endpoint evaluation, 9 tet, 1 tet - overview – Metrohm tiamo 2.2 Manual User Manual
Page 876
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
862
■■■■■■■■
tiamo 2.2
[New]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered.
[Properties]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880).
[Delete]
Deletes the external measured value selected in the table.
5.6.3.8.7.10
STAT U - Fix endpoint evaluation
Dialog window: Method
▶ STAT U ▶ Properties... ▶ Additional evaluations ▶ Fix
endpoint evaluation
▶ [New]/[Properties] ▶ Fix endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-2000.0 ... 2000.0 mV
Time
Range
0.0 ... 999999.9 s
Volume
Range
0.00000 ... 9999.99 mL
5.6.3.9
TET
5.6.3.9.1
TET - Overview
Dialog window: Method
▶ TET ▶ Properties... ▶ TET - 'Command name'
Command for thermometric titration with the 859 Titrotherm.
Devices
This command can be executed with the following devices.
Titrando: 859
Appearance
The command has the following appearance: