2.4 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
36
■■■■■■■■
tiamo 2.2
Identifica-
tion
Description
Commands
.EP.QTY
Measured value for the endpoint in
µg
KFC, BRC
.EP{x}.TEM
Temperature for the endpoint x
(1…9) in °C
DET, MET, SET,
KFT, KFC, BRC
.EP{x}.TIM
Time in s until the endpoint x (1...9)
is reached
DET, MET, SET,
KFT, KFC, BRC
.EP{x}.UGE
Voltage at the generator electrode
for endpoint # (1...9) in V. In case of
measurements with 756, 831: 0 =
not defined, 1 = <14V, 2 =
14...28V, 3 = >28V
KFC, BRC
.EP{x}.VOL
Volume for the endpoint x (1…9) in
mL
DET, MET, SET,
KFT
.EPP
Position after finishing the com-
mand. 0 = invalid position.
PORT
.ETE
End temperature (temperature fol-
lowing processing of the com-
mand) in °C
DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without T/Flow,
Ref, Spec), CAL
MEAS, STDADD,
DOS
.EVT
End volume (total dosed volume at
the end of the command) in mL
DET, MET, SET,
KFT, STAT, DOS
.FIN
Command status; 1 = Command
has ended at least once, 0, inva-
lid (variable not available) = Com-
mand has never ended
All except START
and END
.FP{x}.CNT
Intensity for the fixed endpoint x
(1…9) in counts
MEAS Opt
.FP{x}.DME
Measured value differential for the
fixed endpoint x (1…9)
MET
.FP{x}.DVT
Drift for the fixed endpoint x (1…9)
in µg/min
KFC, BRC
.FP{x}.ERC
ERC for the fixed endpoint x (1…9) DET
.FP{x}.MEA
Measured value for the fixed end-
point x (1...9) in mV
DET, MET, SET,
KFT, STAT