5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
772
■■■■■■■■
tiamo 2.2
Titrando
Range
1 ... 999 µg/min
Default value
20 µg/min
Drift correction
If this parameter is set to automatic or manual the drift correction ( = Drift
value × Time for drift correction) caused by empty consumption during
titration is deducted from the total consumption. The Time for drift cor-
rection is the period between end of conditioning and end of titration.
Selection
automatic | manual | off
Default value
off
automatic
The drift lastly determined during conditioning will be used as drift value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 99.9 µL/min
Default value
0.0 µL/min
Stop volume
Determination stops when the specified volume has been dosed since the
start of the conditioning. The stop volume should be adjusted to the size of
the titration cell in order to prevent any overflow.
Titrando
Range
0.000000 ... 9999.99 mL
Default value
20.00 mL
Selection
off
Default value
off
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Titrando
Range
1 ... 999999 s
Selection
off
Default value
off