5 meas u - additional measured values, Additional measured values – Metrohm tiamo 2.2 Manual User Manual
Page 921
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
907
Window
A range (window) can be defined on the measured value axis or on the time
axis. The break point evaluation will only be carried out in the defined win-
dow. Only the first break point in the defined window will be recognized.
Selection
Measured value | time | off
Default value
off
Lower limit
Value for the lower limit of the window.
Time for the lower limit of the window.
Window = Measured value
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Value for the upper limit of the window.
Window = Measured value
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Window = Time
Range
1 ... 999999 s
Default value
999999 s
5.6.4.4.5
MEAS U - Additional measured values
Tab: Method
▶ MEAS U ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Input
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in the
default settings in additional measured value columns Calculated # or
External # in the measuring point list .