3.11 File manager
86
PC Control / Touch Control
• Current determination data: all the data for the current
determination and the current method with which the determination
was carried out.
• Control: settings under control.
• User list under System settings/User administration and settings
for each user.
• System settings/User administration: all system settings
including device-specific dialog configuration and dialog options for
the list of commands and fixed keys, device-specific settings for the
user administration (login options, password options and audit
trail).
• Titrants: all titrant data.
• Sensors: all sensor data.
• Device data (Touch Control only): all data from the device
manager. It is not possible to transfer the stored device data from
Touch Control to PC Control.
• GLP data: all data from the GLP manager. Please note that the GLP
data cannot be loaded if you load the safety copy on a different
system.
• Common variables: all common variables.
• Templates: all templates for sample data, result templates, remote
lines, custom calibration buffers and report header.
• Routine dialog settings: current routine dialog settings under
Dialog options/Routine dialog.
• Sample Processor rack tables: The racks present in the device
manager of the USB Sample Processor.
• Custom subsequences: Your own subsequences.
• Deactivate the checkboxes for the data blocks which you do not
wish to restore and load the activated data blocks with [Load
data].
You can also load a backup from Touch Control in PC Control and vice
versa. When a Touch Control backup is loaded in PC Control then the
device data is ignored.
After the restoring you must shut down the PC Control software
and then restart it, or switch Touch Control off and then on
again.