3.16.2 Insert command
A method may contain a max. of 99 commands. You can use [Insert
command] to insert a new method command in front of the selected line.
In the example shown a new "empty" method is being edited.
Use [Insert command] to open the dialog Sequence/Insert
command.
The method commands are arranged in different groups. An overview
of all commands and parameters can be found in Section 4. The
individual commands are also described in the online help.
Select the group from which you wish to insert a command. For
Report the command is inserted directly. For all other groups the
command selection list is opened for you to select a command.
After a command has been inserted the list of commands is
automatically shown again. The inserted command is shown as
selected so that you can use [Edit command] directly to edit the
parameters of the newly inserted command (see Section 3.16.1). You
can insert a maximum of nine titrations or measurements and nine
calculations. There is no limit to the number of other commands (up to
99 in one method).