Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 253
4
Parameters
PC Control / Touch Control
241
Open the dialog Edit result/Result limits and activate the
Monitoring result limits
checkbox.
Enter an Upper limit and a Lower limit for the result. If a result is
invalid then it will be treated as if the limits had been infringed.
The following actions are available if the result limits are infringed:
• Display message: The run is halted and a message is shown. You
can choose whether you want to terminate the run or continue with
it. The message will be documented under Results/More deter-
mination data/Messages
.
• Document message: The run is not halted. A message that the result
limits have been infringed will be documented under Results/More
determination data/Messages
.
• Cancel determination: The run is terminated and a corresponding
message will be displayed and documented under Results/More
determination data/Messages
.
The Result variable is assigned automatically when a new result cal-
culation is created. However, it can still be altered later. Only the result
variables Rx (x = 1 to 9), which have not yet been assigned in this
CALC command, can be selected. The value of the result can be ac-
cessed via the result variable. This means that a result can be used for
further calculations.
Under Note you can enter a short text, e. g. to describe the calculation
variables used.