Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 251
4
Parameters
PC Control / Touch Control
239
The result template chosen in the selection list is loaded with [Load
template]
and a note concerning the template is shown. With [Custom
templates]
you can open the list of your own result templates (see
Section 3.14.3,
Section 4.5.4 and Section 4.5.5).
Use [Empty result] to open the edit dialog directly.
• Result name: The standard result name corresponds to the result
variable. The result name is the text which will be shown in the result
display and in the report. It can be altered at any time.
• Calculation formula: The result variable is shown before the cal-
culation formula. The variable can be altered under [Result vari-
able]
. A special editor is opened for the calculation formula if you
touch the input field or click on it (see Section 4.5.3).
• Decimal places: Number of decimals to be shown in the result.
• Result unit: The result unit is shown and saved together with the
result.
Under Result options various parameters for the result calculation
can be edited:
• Variable for mean: If statistical calculations has been switched
on for the method under Sequence/ Method options then the
mean value of the individual results will be stored as the variable
SMNx
(x = 1 to 9). Statistical calculations can be carried out for up to
nine results in one determination. The first free statistics variable will