Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 268
![background image](/manuals/586198/268/background.png)
4.6 Reports
256
PC Control / Touch Control
Report
Parameter
List of measuring
points
- Selection of the titration or measurement (mode)
for which the list of measuring points is to be
printed out.
Calculations
- Determination properties (on/off)
- The variables can be printed out as shown or
with the full accuracy of the floating point num-
ber ("double precision" to Standard IEEE 754).
Devices used
Determination properties (on/off).
Variables
All determination date; depends on command.
Statistics short
Report printout at the end of a statistics series
(statistic counter same as number of samples,
i.e. samples processed = samples set) or at the
end of each determination.
Statistics
Statistics over-
view
Report printout at the end of a statistics series
(statistic counter same as number of samples,
i.e. samples processed = samples set) or at the
end of each determination.
Sample data
Sample data silo
Sample data silo properties, landscape format,
Table with all the sample data that can be en-
tered in the sample data silo.
Result silo
Result silo
Result data silo properties, landscape format,
Table with all the results that can be stored in the
result silo.
Form feed
Form feed
Not editable. Each report is printed out on a
separate page.
No parameters can be edited for the system reports and the method
reports. The contents of the individual reports is described in Sec-
tion 3.23.
Note!
Some tables (statistics overview and sample data silo) are always
printed out in
landscape format, as otherwise the data could not be
printed out on a single page. You must select
portrait in the Win-
dows-specific
Printer setting nevertheless, as printout in landscape
format is automatic.