System file (touch control only) – Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 312

5.2 Diagnosis
300
PC Control / Touch Control
PC Control:
When the update has been completed you must shut down the PC
Control software. This is done by pressing [Back] or [Home] and
confirming the following message with the request for a restart with
[OK]. The PC Control program ends automatically.
Restart PC Control.
Touch Control:
When the update has been finished you must switch off the Touch
Control, wait a few seconds, and then switch it on again.
Update of USB Sample Processor (PC Control and Touch
Control)
Select the USB Sample Processor to be updated in the menu
System/Diagnosis/Software update under Device and press
[Continue].
[File] accesses the selection of the program files. The file name
has the following structure:
USB Sample Processor – program:
5xxx00yy.BIN
xxx = USB Sample Processor (device number)
yy = 10 or higher (program version)
Take care that the device and version numbers are correctly
assigned.
Select the required program file with [Open].
Start the update with the [Start] fixed key. Further steps are
shown in the dialog.
If the USB Sample Processor is updated from the Touch Control
then the menu guidance is slightly different from that described
above.
PC Control:
When the update has been completed you must shut down the PC
Control software. This is done by pressing [Back] or [Home] and
confirming the following message with the request for a restart with
[OK]. The PC Control program ends automatically.
Restart PC Control.
Touch Control:
When the update has been finished you must switch off the Touch
Control, wait a few seconds, and then switch it on again.
System file (Touch Control only)
If you want to install new devices, e. g. balances or printers, for
which there are no drivers in the control software you can load them
by using this function. The procedure is identical to the update of
the Touch Control.