4 input lines – Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 112

3.14 Templates
100
PC Control / Touch Control
3.14.4 Input lines
Binary patterns can be defined for scanning remote lines with the
SCAN command (see Section 4.11.1) during the determination
sequence; these can then be selected when editing the parameters.
Open the dialog Templates/Input lines.
You can define new templates with [New]. Existing templates can be
removed from the list with [Delete] or edited with [Edit].
List of predefined input lines
Parameter Signal
pattern
Function
Cond OK
******1*
waits for "Cond OK" condition of device 1
End1
****1***
waits for the EOD impulse from device 1 (Titrino,
Titrando)
End2
*1******
waits for the EOD impulse from device 2
EndMeter
***11***
waits for the EOD impulse from ion meter or pH-Meter
(stirrer 1 will be switched on during the waiting period)
Ready*
**1****1
waits for "ready" condition of devices 1 and 2 (Titrino,
Titrando)
Ready1
*******1
waits for "ready" condition of device 1
Ready2
**1*****
waits for "ready" condition of device 2
Sample ready
***1****
waits for an impulse to continue e. g. of a connected
Sample Processor as soon as it is ready
With the parameter Ready* it is possible to scan the condition of
devices working in parallel. In this case the 'Ready' line of both devices
must be statically set (i.e. constantly) at the end of a determination.
Devices which only transmit a brief impulse at the end of e. g. a
measurement cannot be controlled in parallel.
Use [Edit] to open the edit dialog for the template Cond OK.