beautypg.com

Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 269

background image

4

Parameters

PC Control / Touch Control

257

4.7 Calibrating pH electrodes (CAL pH) and ISE

(CAL Conc)

For pH measurements, titrations to a preset endpoint (SET pH) and

the determination of pK values and fixed endpoints the calibration of

the pH electrode (command CAL pH) is essential; it is also recom-

mended for all other pH titrations. As least two buffers should be used

for the calibration; these should cover the expected measuring range.

For concentration direct measurements with ISE electrodes the sen-

sor must be calibrated with a series of standards (CAL Conc). During

the sequence the calibration data is stored automatically together with

the sensor data for the used sensor.
Under Calibration parameters you can edit the parameters for the

calibration sequence. These are identical for pH calibrations and ISE

calibrations. In the command CAL pH you must define the calibration

buffers, in the command CAL Conc the concentrations of the stan-

dards.
Signal drift (d): Drift, i.e. the alteration in the measured value per

minute that has to be reached for measured value acceptance. off

means that the measured value acceptance takes place when the

maximum waiting time has elapsed. This may be advisable if the

electrode has a very slow response.

Waiting time min. (d): The measured value is not accepted until

the minimum waiting time has elapsed, even if the measured value

drift has already been reached. The minimum waiting time is only

important for drift-controlled measurements.

Waiting time max. (d): If the signal drift is switched off or has not

yet been reached then the measured value will be accepted after

the maximum waiting time has elapsed. If the waiting time has not

been newly entered then a waiting time that is suitable for the drift

will be calculated automatically according to the following equation:

5

0.01

Drift

150/

time

Waiting

+

+

=

Temperature (d): Manually entered calibration temperature. If a

temperature sensor is connected and temperature measurement is

set to automatic or continuous under Sensor (see Section 4.1.5)

then the temperature will be measured. If off is defined then the

temperature is requested after the start (except the calibration is

carried out with a Sample Processor).

Sample Processor (d): If the calibration is carried out with a Sam-

ple Processor then the buffers or standards are changed automati-

cally. That is why there is no request of the calibration temperature

at the start of the calibration. The value entered above is used.
For calibrations with a Sample Processor connected via a remote

box the parameter Remote must be activated. When the measure-

ment of a buffer or standard has been completed the Titrando

sends a signal to the Sample Processor via the remote connection

(EOD, see Section 6.2.2). The Sample Processor then switches to

the next buffer or standard.