5 save determination, 6 load determination data – Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 148

3.18 Results and more determination data
136
PC Control / Touch Control
determination are deleted and you return to the properties of the
determination.
3.18.5 Save determination
You can save the current determination on Card 1, Card 2 or a
shared file system (requires 847 USB Lab Link, see Section 3.10.11). In
this way you can store all the data of a determination. If you wish to
save every determination then you can use the function Save
automatically (see Section 3.16.10). Under Sequence/Method
options you can define that all the determination data is to be saved at
the end of each determination. If the data is saved automatically with
the standard settings then each determination will be automatically
write-protected. In this way your original data is protected and can
neither be edited nor deleted.
In the dialog Results/More determination data, open the
selection Load/Save.
Press Save. The procedure for saving a file is described in
Section 3.11.4. The file name Identification 1-YYYYMMDD-hhmmss
is proposed as default. This means that the file name consists of
the text entered as Identification 1 and the determination time. You
can also enter any other file name you choose. The extension of the
file name with the determination time has the advantage that the file
names of all determinations are unambiguous.
The following data is stored for each determination:
• Calculated results.
• All raw data (measuring point lists, endpoints...) and variables
produced during the determination.
• Determination properties.
• Statistics, if statistical calculations have been carried out.
• Method with which the determination was carried out.
• Common variables (values and properties of all common variables
at the time of the determination).
• Most important titrant data for the titrants used.
• Most important calibration data for the sensors used.
Note!
The data saved with each determination ensures that the data is
traceable and that the documentation is GLP-conform.
3.18.6 Load determination data
You can load determination files from Card 1, Card 2 or a shared file
system (requires 847 USB Lab Link, see Section 3.10.11).
In the dialog Results/More determination data open the
selection Load/Save. Touch Control: If no card is inserted or no
external file system is shared, then the [Load] button is inactive.
Open the dialog Load and select [Card 1], [Card 2] or [Shared
memory] (if present).