4

Parameters

PC Control / Touch Control

279

wish to use a pulse with a pulse length other than 200 ms then you

must define a corresponding template.

We recommend that non-relevant output lines are also masked with an

asterisk (*) in order that the condition of these lines is not altered.
The templates are described in Section 3.14.5.

4.11.3 Receive data (SCAN RS)

Definition of the parameters for receiving data via a serial RS-232 inter-

face.
Serial port: Selection of the RS-232 interface which the peripheral

device is connected to.

Data string: Enter the desired data string here. "*" can be used for

one or more characters.

Timeout: You can define a time after which an action should be car-

ried out in case the signal pattern has not been recognized.

Timeout action: The action defined here will be carried out if the

time interval has expired.

4.11.4 Send data (CONTROL RS)

Definition of the parameters for sending data via a serial RS-232 inter-

face.
Serial port: Selection of the RS-232 interface which the peripheral

device is connected to.

Data string: Enter the desired data string here. All characters ac-

cording to ASCII code can be used. Please consult the online help

for how to enter the data string.

Note!

Each command is terminated automatically with carriage return + line

feed.