beautypg.com

Limit values of the three electrode types – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 280

background image

4.9 Electrode test for pH electrodes (ELT)

268

PC Control / Touch Control

Limit values of the three electrode types

Rating

Excellent electrode

Good electrode

Usable electrode

Electrode type "Standard"

Streaming potential

2.5 mV

3.0 mV

4.0 mV

Drift

2.0 mV/s

2.5 mV/s

3.0 mV/s

Slope min.

96.5 %

96.0 %

95.0 %

Slope max.

101.0 %

102.0 %

103.0 %

Response time

45 s

50 s

60 s

Lower limit Uoff

–15 mV

Upper limit Uoff

15 mV

Electrode type "Gel"

Streaming potential

3.0 mV

3.5 mV

4.5 mV

Drift

2.5 mV/s

3.0 mV/s

4.0 mV/s

Slope min.

96.5 %

96.0 %

95.0 %

Slope max.

101.0 %

102.0 %

103.0 %

Response time

60 s

75 s

90 s

Lower limit Uoff

–15 mV

Upper limit Uoff

15 mV

Electrode type "Non-aqueous"

Streaming potential

3.0 mV

4.5 mV

6.0 mV

Drift

5.0 mV/s

7.0 mV/s

9.0 mV/s

Slope min.

88.0 %

80.0 %

70.0 %

Slope max.

120.0 %

130.0 %

140.0 %

Response time

60 s

75 s

90 s

Lower limit Uoff

–10 mV

Upper limit Uoff

70 mV