3
Operation
PC Control / Touch Control
69
Activate the checkbox Spooler in order that the Touch Control is
not blocked by printing out a report so that you can continue
working.
Note!
The Touch Control prints the reports with a fixed resolution of 300 dpi.
If you are using a printer with a resolution of 360 dpi (or a multiple,
e. g. Epson) then the text will be printed out somewhat smaller than
with printers with a resolution of 300 dpi (or a multiple, e. g. Canon or
HP).
3.10.8 Balance
Connecting a balance to the Titrando or PC is described in the
Installation Instructions for the Titrando. In the device manager you
must configure the serial interface used to connect the balance.
Select the Balance type.
The following table shows the balance type to be selected for various
balance models:
Balance
Balance type
AND
AND
Mettler AB, AG, AM, PM, XP, XS
Mettler
Mettler AT
Mettler AT
Mettler AX, MX, UMX, PG, AB-S
Mettler AX
Ohaus
Voyager, Explorer, Analytical Plus
Ohaus
Precisa
Precisa
Sartorius
Sartorius
Shimadzu BX, BW
Shimadzu
PC Control:
Under RS-232 select the RS-232 interface (COM) of the PC to which
the balance is connected. All the COM interfaces of the PC will be
shown.
Add, if not present already, the virtual device Serial ports to the
device manager and open the list with the available RS-232
interfaces with [Edit]. Select the interface to which the balance is
connected. Open the dialog with the port parameters with [Edit]
and configure these. The settings on the balance must agree with
these settings here. A description of the parameters can be found in
the online help.