3 glp tests for measurement and titration – Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 101
3
Operation
PC Control / Touch Control
89
3.12.3 GLP tests for measurement and titration
You can document the GLP tests for measurement and titration
here.
Open the dialog GLP manager/GLP tests HW/SW (HW = hardware,
SW = software).
Select [GLP test measurement] or [GLP test titration]
according to which GLP test you wish to document.
For both tests you can select or enter the Method with which the test
is to be carried out. Only methods in the internal memory will be
shown in the selection list. You can also enter or select the User
who carried out the test from the selection list.
Enter the Test date for the last test and select the Test result.
You can enter a Comment for each test.
Press [Hardware].
Select the Test tool (list under GLP manager/Test tools), Control
device (configured in Device manager), Measuring input and
Dosing device (GLP test titration only) used for the test.
You can monitor the time interval after which a GLP test must be
carried out again. (A detailed description of the monitoring function
using "Titer validity" as an example can be found in Section 3.8.4.)