beautypg.com

3 glp tests for measurement and titration – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 101

background image

3

Operation

PC Control / Touch Control

89

3.12.3 GLP tests for measurement and titration

You can document the GLP tests for measurement and titration

here.

Open the dialog GLP manager/GLP tests HW/SW (HW = hardware,

SW = software).

Select [GLP test measurement] or [GLP test titration]

according to which GLP test you wish to document.

For both tests you can select or enter the Method with which the test

is to be carried out. Only methods in the internal memory will be

shown in the selection list. You can also enter or select the User

who carried out the test from the selection list.

Enter the Test date for the last test and select the Test result.

You can enter a Comment for each test.

Press [Hardware].

Select the Test tool (list under GLP manager/Test tools), Control

device (configured in Device manager), Measuring input and

Dosing device (GLP test titration only) used for the test.

You can monitor the time interval after which a GLP test must be

carried out again. (A detailed description of the monitoring function

using "Titer validity" as an example can be found in Section 3.8.4.)