Flow rate of a single dosing device, Filling times – Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 287
4
Parameters
PC Control / Touch Control
275
Flow rate of a single dosing device
The cylinder volume and the filling time of the exchange or dosing
unit must be taken into account for the calculation of the effective
flow rate.
Max. flow rates for different cylinder volumes:
Cylinder volume
Max. flow rate
Exchange unit
Dosing unit
1 mL
approx. 80 mL/h
---
2 mL
---
approx. 170 mL/h
5 mL
approx. 400 mL/h
approx. 430 mL/h
10 mL
approx. 800 mL/h
approx. 860 mL/h
20 mL
approx. 1.6 L/h
approx. 1.7 L/h
50 mL
approx. 4 L/h
approx. 4.3 L/h
Filling times
Filling times are not included in the calculation of the dosing rate by
the Titrando. The filling times (including valve switching) can be es-
timated according to the following formula:
4s
t
rate
filling
Actual
rate
filling
Maximum
s
in
time
Filling
+
⋅
=
t = 20 s (for exchange units)
18 s (for dosing units)
The maximum filling rate depends on the cylinder volume of the ex-
change or dosing unit used (see Section 3.8.6). In the table below
the approximate filling times (including valve switching) are given for
various cylinder volumes (exchange and dosing unit) at different fill-
ing rates:
Cylinder volume
Filling time in s
maximum
100 mL/min
50 mL/min
10 mL/min
1 mL
24 / ---
---
---
---
2 mL
--- / 22
---
---
---
5 mL
24 / 22
---
---
34
10 mL
24 / 22
---
---
64
20 mL
24 / 22
---
28
124
50 mL
24 / 22
34
64
304