Flow rate of a single dosing device, Filling times – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 287

background image

4

Parameters

PC Control / Touch Control

275

Flow rate of a single dosing device
The cylinder volume and the filling time of the exchange or dosing

unit must be taken into account for the calculation of the effective

flow rate.
Max. flow rates for different cylinder volumes:

Cylinder volume

Max. flow rate

Exchange unit

Dosing unit

1 mL

approx. 80 mL/h

---

2 mL

---

approx. 170 mL/h

5 mL

approx. 400 mL/h

approx. 430 mL/h

10 mL

approx. 800 mL/h

approx. 860 mL/h

20 mL

approx. 1.6 L/h

approx. 1.7 L/h

50 mL

approx. 4 L/h

approx. 4.3 L/h

Filling times
Filling times are not included in the calculation of the dosing rate by

the Titrando. The filling times (including valve switching) can be es-

timated according to the following formula:

4s

t

rate

filling

Actual

rate

filling

Maximum

s

in

time

Filling

+

=

t = 20 s (for exchange units)

18 s (for dosing units)

The maximum filling rate depends on the cylinder volume of the ex-

change or dosing unit used (see Section 3.8.6). In the table below

the approximate filling times (including valve switching) are given for

various cylinder volumes (exchange and dosing unit) at different fill-

ing rates:

Cylinder volume

Filling time in s

maximum

100 mL/min

50 mL/min

10 mL/min

1 mL

24 / ---

---

---

---

2 mL

--- / 22

---

---

---

5 mL

24 / 22

---

---

34

10 mL

24 / 22

---

---

64

20 mL

24 / 22

---

28

124

50 mL

24 / 22

34

64

304