4.5 Calculations
250
PC Control / Touch Control
Variable Description
Method command
MIT
Initial temperature (temperature before
start conditions are processed)
DET, MET, SET,
KFT, STAT, MEAS,
DOS
MSA
Start volume which is added according to
the start volume given in the start condi-
tions.
DET, MET
MSP
Start volume which is added according to
the start measured value given in the start
conditions.
DET, MET
MSS
Start volume which is added according to
the start slope given in the start conditions.
DET, MET
MSV
Start volume for all start conditions (start
volume, start measured value and start
slope)
DET, MET, SET,
KFT, STAT
MSD
Time for processing the start conditions
DET, MET, SET,
KFT, STAT
MSM
Start measured value (measured value af-
ter processing the start conditions)
DET, MET, SET,
KFT, STAT, DOS
MST
Start temperature (temperature after proc-
essing the start conditions)
DET, MET, SET,
KFT, STAT, DOS
MCV
End volume (total volume added at end of
titration)
DET, MET, SET,
KFT, STAT, MEAS
Conc, DOS
MCD
Total duration of titration, measurement or
calibration (including start conditions)
DET, MET, SET,
KFT, STAT, MEAS,
DOS, CAL
MTM
Method of temperature measurement
(Pt1000, NTC, manual); format = text
DET, MET, SET,
KFT, STAT, MEAS,
DOS
MDD
Duration of net dosing (without filling
times, pause)
DOS
MDC
Drift for drift correction
SET, KFT
DDC
Time for drift correction
SET, KFT
MCM
Final measured value (measured value at
end of the titration or measurement);
MEAS: corresponds to the measured con-
tent
DET, MET, SET,
KFT, STAT, MEAS,
DOS
MCT
Final temperature (temperature at end of
the titration, measurement or calibration)
DET, MET, SET,
KFT, STAT, MEAS,
DOS, CAL