4 evaluation – Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 243

4
Parameters
PC Control / Touch Control
231
4.4 Evaluation
A list of measuring points can have various additional evaluations
applied to it. The evaluation (EVAL command) always refers to the last
titration or measurement in the method sequence. Which evaluations
can be used for which titrations and measurements is given in the fol-
lowing table:
Evaluation
Titration
Measurement
EVAL Fix-EP
Fixed end-
points
EVAL pK/HNP
pK value
Half neutraliza-
tion potential
EVAL MIN/MAX
Minimum/
Maximum
EVAL BREAK
Break points
EVAL RATE
Rate evaluation
DET pH
V, MV, t
pK value
yes
yes
–
DET U
V, MV, t
HNP
yes
yes
–
DET Ipol
V, MV, t
–
yes
yes
–
DET Upol
V, MV, t
–
yes
yes
–
MET pH
V, MV, t
pK value
yes
yes
–
MET U
V, MV, t
HNP
yes
yes
–
MET Ipol
V, MV, t
–
yes
yes
–
MET Upol
V, MV, t
–
yes
yes
–
SET pH
V, MV, t
–
yes
–
yes
SET U
V, MV, t
–
yes
–
yes
SET Ipol
V, MV, t
–
yes
–
yes
SET Upol
V, MV, t
–
yes
–
yes
KFT Ipol
V, MV, t
–
yes
–
yes
KFT Upol
V, MV, t
–
yes
–
yes
STAT pH
V, MV, t
–
yes
–
yes
STAT U
V, MV, t
–
yes
–
yes
MEAS pH
MV, t
–
yes
yes
–
MEAS U
MV, t
–
yes
yes
–
MEAS T
MV, t
–
yes
yes
–
MEAS Ipol
MV, t
–
yes
yes
–
MEAS Upol
MV, t
–
yes
yes
–
MEAS Conc
– – – – –
MEAS Cond
MV, t
–
yes
yes
–
Only those evaluations which are available for the last titration or meas-
urement before the EVAL command can be inserted in the list of com-
mands. If a titration or measurement is deleted before the EVAL com-