1 evaluation of fixed endpoints (eval fix ep) – Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 244

4.4 Evaluation
232
PC Control / Touch Control
mand then it will be shown in red in the list of commands, as the refer-
ence is missing.
You can also insert EVAL commands in the method sequence at a later
date and trigger the re-evaluation under Results with [Recalculate].
The data for each evaluation can be shown under Results/More de-
termination data/View data
and can be calculated in CALC com-
mands.
4.4.1 Evaluation
of
fixed endpoints (EVAL FIX EP)
Fixed endpoints can be determined for all titrations and measurements
(except MEAS Conc). For a fixed quantity (measured value, volume or
time) the associated values are interpolated from the list of measuring
points.
• Fixed quantity: Selection of the fixed quantity.
• Fix EP1 at to Fix EP9 at: For the entered measured value, vol-
ume or time the associated values for the other quantities are inter-
polated from the list of measuring points. The fixed EP must lie be-
tween the first and last entries in the list of measuring points.
Up to nine fixed endpoints can be evaluated for each EVAL FIX EP.