3.16 Editing parameters

120

PC Control / Touch Control

Only Identification 1 is shown with the title "Charge". The fixed sample

size "10 mL" is shown and cannot be edited.
You can also define the limits for the sample size which will then be

monitored in the determination sequence.

Open the dialog Sample data/Sample size limits and activate

the Monitoring sample size limits checkbox.

Enter a Lower limit and an Upper limit for the sample size.

The limits will not be monitored when the sample size is entered in the

main dialog, but at the START of the determination and at its end, when

an automatic sample data request is made with a REQUEST command

and on recalculating a determination. It does not matter whether the

sample size has been entered in the main dialog "by hand", transmitted

by a balance or barcode reader or entered in the sample data silo.

When monitoring is switched on and the limits for the sample size are

infringed then an entry in the message list for the determination will

be made automatically (see Section 3.18.2).
Activate the Display message checkbox if a message is to be shown

during a determination sequence to indicate that the sample size limits

have been infringed. You then have the choice of continuing the run or

of canceling it. This Message can be provided with a Time out, i.e. it

will vanish automatically when the given time has elapsed and the run

will be continued automatically.