beautypg.com

Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 65

background image

3

Operation

PC Control / Touch Control

53

Under Calibration data/History the calibration data, dates and

times of the last ten calibrations are shown in the form of a table.

Calibration data which have been determined automatically is shown in

green, manually entered calibration data is shown in black and marked

with (m). With [Graph slope] and [Graph pH(0)], or [Graph E(0)]

you can open a graph in which the slope or pH(0) for pH electrodes or

E(0) for ISE electrodes is plotted against the calibration date. If a sensor

has to be replaced then you can delete the complete history with

[Delete history].

This graph corresponds to a control chart on which both warning

limits and intervention limits are shown for the calibration data. The

values for these warning limits and intervention limits can be entered

under Graph/Limits. No automatic action is taken when these limits

are infringed.

3.9.5 Monitoring the calibration interval (for pH and ISE electrodes

only)

You can monitor the time interval after which the sensor must be

recalibrated. (A detailed description of the monitoring function using

"Titer validity" as an example can be found in Section 3.8.4.) For

documentation purposes you can enter the starting date without

monitoring the time interval.