4.12 Automation
284
PC Control / Touch Control
The Conditioning sequence will be carried out just before the
conditioning. It gives you the opportunity, for example, to start a
connected Polytron
®
high-frequency homogenizer and to set its stir-
ring rate before the start of the conditioning. Use the CONTROL RS
command for this.
• Cycles: A subsequence is repeated in accordance with the number
of runs. "Calibration" is required if automatic calibration is to be
carried out with a USB Sample Processor. Apart from the CAL
command, the subsequence contains a MOVE command to move
to the calibration solutions. It is processed until the number of buff-
ers or standards defined in the calibration command has been
reached.
In a [Note] a short text can be used to describe e. g. what the subse-
quence is used for.
With [Insert command] you can insert a command in front of the se-
lected line.
Note!
In contrast to general method programming, the choice of commands
is limited. With the exception of calibration, no commands for re-
cording or evaluating measured values (e. g. titration, measurement)
can be inserted in a subsequence. Only one CAL command can be
inserted in a subsequence.
[Delete command]
removes the selected command from the list. The
parameters for the selected command can be edited with [Edit com-
mand]
.