Metrohm 840 PC Control 5.0 / Touch Control User Manual
Page 114
![background image](/manuals/586198/114/background.png)
3.14 Templates
102
PC Control / Touch Control
METER Mode C
*********1000*
switches ion meter to Conc measurement
METER Mode I
*********0100*
switches ion meter or pH meter to Ipol (mV
measurement) and starts it
METER Mode pH
*********0001*
switches ion meter or pH meter to pH measurement and
starts it
METER Mode T
*********0010*
switches ion meter or pH meter to temperature
measurement and starts it
METER Mode U
*********0011*
switches ion meter or pH meter to mV measurement and
starts it
Start device1
*************p
starts device 1(e. g. Titrino, Titrando, ...) *
)
Start device2
********p*****
starts device 2
(see above, only with multiple cable)*
)
Start device*
********p****p
starts devices 1 and 2 " *
)
Start Dos1
*******p******
starts Dosimat at device 1 (Titrino via "activate")
Start Dos2
*****p********
starts Dosimat at device 2 "
Start Dos*
*****p*p******
starts Dosimat at devices 1 and 2 "
For
START
commands the signal is transmitted as a short impulse of 200 ms.
*
)
With pH meters or ion meters a result report is triggered
Use [Edit] to open the edit dialog for the template Continue.
You can also set pulses for the lines by entering a p in the binary
pattern. The pulse length is defined under Pulse length for "p". The
pulse length can only be entered for signals which are defined as
templates. If you define a signal directly in the CTRL command or in
manual operation then a fixed pulse length of 200 ms will be used. To
input the binary pattern please consult the online help.