3.14 Templates

102

PC Control / Touch Control

METER Mode C

*********1000*

switches ion meter to Conc measurement

METER Mode I

*********0100*

switches ion meter or pH meter to Ipol (mV

measurement) and starts it

METER Mode pH

*********0001*

switches ion meter or pH meter to pH measurement and

starts it

METER Mode T

*********0010*

switches ion meter or pH meter to temperature

measurement and starts it

METER Mode U

*********0011*

switches ion meter or pH meter to mV measurement and

starts it

Start device1

*************p

starts device 1(e. g. Titrino, Titrando, ...) *

)

Start device2

********p*****

starts device 2

(see above, only with multiple cable)*

)

Start device*

********p****p

starts devices 1 and 2 " *

)

Start Dos1

*******p******

starts Dosimat at device 1 (Titrino via "activate")

Start Dos2

*****p********

starts Dosimat at device 2 "

Start Dos*

*****p*p******

starts Dosimat at devices 1 and 2 "

For

START

commands the signal is transmitted as a short impulse of 200 ms.


*

)

With pH meters or ion meters a result report is triggered

Use [Edit] to open the edit dialog for the template Continue.

You can also set pulses for the lines by entering a p in the binary

pattern. The pulse length is defined under Pulse length for "p". The

pulse length can only be entered for signals which are defined as

templates. If you define a signal directly in the CTRL command or in

manual operation then a fixed pulse length of 200 ms will be used. To

input the binary pattern please consult the online help.