beautypg.com

Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 295

background image

4

Parameters

PC Control / Touch Control

283

The list with templates for subsequences is displayed under Edit Sub-

sequence/Load

:

Open the list with your [Own subsequences]. You have now the possi-

bility to load or to delete your own subsequence.
Under [Subsequence options] parameters are set that refer to the

whole subsequence:

Subsequence type: The type of subsequence is entered here.

A Sample sequence is carried out each time that a method is run.
The Start sequence is only carried out at the start of a sample se-

ries when the autostart counter = 1.
If a subsequence is only to be carried out for the last sample in a

series, i.e. when the autostart counter has reached its set value,

then it is defined as the End sequence.
The Stop sequence will only be carried out when a method is termi-

nated because of the following cases:
- Manual termination with the [STOP] fixed key
- Termination because of an error
- Termination by remote signal via the Control remote box
If the method is ended regularly then the stop sequence is not car-

ried out. The sample sequence can be used in this case.