beautypg.com

3 measurements (meas cond) – Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 242

background image

4.3 Measurements (MEAS Cond)

230

PC Control / Touch Control

4.3 Measurements (MEAS Cond)

The parameters for the measurement sequence are edited under Meas-

uring parameters

.

Signal drift (d): Drift, i.e. the alteration in the measured value per

minute that has to be reached for measured value acceptance. off

means that the measured value acceptance takes place when the

maximum waiting time has elapsed. This may be advisable if the

electrode has a very slow response.

Note!

A

constant measured value is often only achieved after a certain

time, as mixing and possibly the reaction itself require a certain time.

The response time of an electrode can also increase with time, i.e.

reaching a constant measured value takes longer and longer. In such

a case

drift-controlled measured value acceptance is particularly

advisable, as the measured values are only accepted when equilib-

rium has almost been achieved.

Waiting time min. (d): The measured value is not accepted until

the minimum waiting time has elapsed, even if the measured value

drift has already been reached. The minimum waiting time is only

important for drift-controlled measurements.

Waiting time max. (d): If the signal drift is switched off or has not

yet been reached then the measured value will be accepted after

the maximum waiting time has elapsed.

Stop measured value (d): Stops when the preset measured value

has been reached since the start of the measurement.

Time interval MP (d): Time interval for entering a measuring point

in the list of measuring points. This list can contain max. 1000

measuring points.

Editing the parameters for the temperature compensation:
Measuring temp. (d): Manually entered measuring temperature. If a

temperature sensor is connected and temperature measurement is

set to automatic or continuous under Sensor (see Section 4.1.5)

then the temperature will be measured continuously.

Reference temp. (d): The measured conductivity is converted to

this temperature.

Temp. coefficient (d): The temperature coefficient is required for

converting the measured conductivity to the reference temperature.

The parameters for Control device, Sensor and Stirrer are de-

scribed in Section 4.1.5. The selection of the Direct parameters is car-

ried out as described in Section 4.1.6.