Page 40

background image

3.7 System settings

28

PC Control / Touch Control

status "Saved" can be signed. The status of the method is given under

Method options/Properties (see Section 3.16.9).

If use only released methods is activated then the corresponding

user can only use methods that have been released, i.e. can only start

methods that have been signed at level 2. Review methods

(signature level 1) must be activated for a user to be entitled to

sign a method at level 1. The method then receives the status

"Reviewed". To be able to sign methods at level 2 Release methods

(signature level 2) must be activated. A method can only be signed

at level 2 when it has already been signed at level 1. The same user can

sign different methods either at level 1 or at level 2, if both checkboxes

Review methods and Release Methods are activated but not the same

method at the same time at level 1 and at level 2. When the user has

signed the method at level 2 it receives the status "Released". If Delete

signatures has been activated for a user then the user is entitled to

delete signatures. However, this is possible for methods that have been

signed at level 2. All the signatures are deleted and the method

receives the status "Saved".
The methods can be signed under Method options/Properties. The

procedure is described in Section 3.16.3.
By clicking on [Signature determination] you can access the edit

dialog for assigning the rights for using and signing determinations.

These settings are only effective when login and password is used.