3.19 Sample data
148
PC Control / Touch Control
You should now enter the sample data, irrespective of whether you are
using PC Control or Touch Control.
Under Method you can select the name of the method with which
the sample is to be processed. You can use Sample data
silo/Properties to define the Method memory (internal memory,
Card 1, Card 2, Shared memory) from which the method is to be
loaded. You can also enter the method name. If no method is
entered then the current method will be used. If you enter the same
method in successive silo lines then the method will not be loaded
separately for each determination.
If you have switched on the use of the Sample assignment table
under System/Templates/Sample data then you cannot define a
method in the sample data silo. If a method has already been entered
before the sample assignment table is activated then this entry will be
ignored.
You can enter sample-specific data under Identification 1 and
Identification 2.
The sample identifications can be accounted as the variables CI1 and
CI2 (see Section 4.5.7). The labels of the sample identifications can be
edited specifically for the method (see p. 150). Under
System/Templates/Sample data you can create a sample
identification list. You can then make a selection from this template
when entering the identifications. This simplifies the input of frequently
used texts for the identification. You can also enter the identifications
with the aid of a barcode reader or send them from a connected
balance. With Touch Control we recommend the connection of a PC
keyboard if long texts are to be entered as identifications.
Enter the value and unit for the sample size. If you have retained
the standard settings under Sample data silo/Properties then
the silo line will be concluded with the sample size. With PC Control
if you confirm the input of the value with [Enter] then a new silo
line will be generated automatically. With Touch Control if you
confirm the input of the value with [OK] then the program will
automatically change to the edit dialog for the next silo line.
The value for the sample size can be calculated as the variable C00
(see Section 4.5.7). The sample size unit can either be selected from a